UV Lamp Aging Test Tester

The UV aging test chamber is suitable for all kinds of geo-synthetics, such as light resistance, weather resistance and light ageing
The exposure test of the sample was carried out in a fluorescent UV lamp vessel in a controlled environment. It can simul...

Leave a MessageService Hotline
+86(21)-64208466

Product Advantages

  1. Cost-Efficient Performance: Low operational and maintenance costs, paired with intuitive operation for seamless integration into any lab.
  2. Advanced Automation: Programmable controller enables unattended test cycles with precise timing and automatic transitions between light, condensation, and dark phases.
  3. Consistent & Repeatable Results: Stable UV output and tight temperature uniformity (±3℃) guarantee high repeatability across tests.
  4. Versatile Simulation: Replicates harsh natural climates—UV radiation, humidity, condensation, and thermal cycling—to accelerate aging processes.
  5. Customizable Testing: Optional UV-A340 (315–400nm) or UV-B313 (280–315nm) lamps cater to diverse material requirements.

Product Features

  • Multi-Factor Aging Simulation: Combines UV exposure, high-temperature aging, and condensation cycles to evaluate color fading, strength loss, cracking, oxidation, and adhesion failure.
  • Intelligent Cycle Control: Programmable light/condensation cycles (e.g., 4H/4H or 8H/8H) with a 0–9999h59min timer.
  • Real-Time Monitoring: Digital displays track temperature, humidity, and irradiation intensity during testing.
  • Compact Design: Optimized chamber layout ensures uniform UV distribution and efficient space utilization.
  • Wide Applicability: Ideal for geosynthetics, automotive coatings, plastics, textiles, and rubber products.

 

Standard 

  • JTG E50 T1164: Testing methods for geosynthetics in highway engineering.
  • GB/T 14522-93: Accelerated weathering tests for plastics, coatings, and rubber in industrial machinery.
  • GB-T16422.3-1997 & ISO4892-3:1994: Laboratory light source exposure using fluorescent UV lamps.

Technical Parameters

Parameter Specification
Light Temperature Range RT +10℃ to 70℃
Condensation Temperature 50℃ (steam humidification)
Temperature Uniformity ±3℃
Irradiation Intensity 0.55 W/m² (at 340nm)
UV Wavelength Options UVA (315–400nm) or UVB (280–315nm)
Light/Condensation Cycle 0–9999h59min (adjustable)
Lamp Center Distance 70mm
Sample-to-Lamp Distance 55mm
Control System Microprocessor PID with overheat protection



 

 