Structure and Materials:
- Working Chamber Dimensions: 660×1150×600mm (Depth×Width×Height)
- Overall Dimensions: Approx. 780×1380×1780mm
- Inner Chamber Material: SUS304 stainless steel plate
Exterior Material: SUS304 stainless steel plate - UV Lamps: UVA-340 fluorescent UV lamps (8 units, 40W each)
- Water Supply & Consumption: 8 liters/day of purified or distilled water (user-provided)
- Control System: 7-inch touch screen, PLC programmable controller
Test Equipment:
The chamber is constructed of corrosion-resistant metal materials, equipped with 8 fluorescent UV lamps, sample racks, and temperature/time control systems with indicators.
- Lamp Power: 40W per lamp.
- Lamp Arrangement: Lamps are installed in a single row, with two groups of four lamps each controlled by separate switches.
- Sample Installation: Test samples are fixed at a distance of 50mm from the nearest lamp surface.
Test Conditions:
- Sample Placement: Test samples are mounted on sample racks.
- Test Temperature: Adjustable between RT+10~70℃ during light exposure, with a temperature tolerance of ±3℃.
- Light/Condensation Cycles: Options include 4H light + 4H condensation, 8H light + 4H condensation, or other custom cycles.
- Lamp Replacement: Replace 1-2 lamps after the first 400–450 hours of use. Subsequent replacements follow a rotational schedule to compensate for lamp aging, ensuring a service life of 1,600–2,000 hours per lamp.
Sample and Performance Evaluation:
- Visual Assessment:
- Primarily for coatings (gloss, color change, chalking, spotting, blistering, cracking, dimensional stability).
- Plastics and rubbers may also be evaluated visually if required.
- Quantitative measurements are recommended using instruments (e.g., gloss meter, colorimeter).
- Mechanical and Other Properties:
- Rubber: Tensile strength, elongation, hardness.
- Plastics: Impact strength, elongation at break, tensile strength, flexural strength.
- Additional tests may be specified as needed.