A Carbon Black Content Tester is an instrument used to measure the carbon black content in materials.
Applications
The Carbon Black Content Tester has wide applications in industrial production, scientific research, and quality control. In industrial production, it can be used to quantitatively analyze the carbon black content in raw materials, thereby controlling the raw material ratio and product quality during the production process. In scientific research, it can be used to quantitatively analyze the carbon black content in new materials, thus understanding the physical and chemical properties of the materials. In quality control, it can be used to quantitatively analyze the carbon black content in products, thereby ensuring the quality and safety of the products.
Standards
GB 13021-1991: Determination of carbon black content in polyethylene compounds.
GB/T 2951.41-2008: Test method for carbon black and ash content in cables and wires.
International Adaptability: Configurable for ASTM D1603 and ISO 6964 equivalency upon request.
Features
Pyrolysis System:
Split-Tube Furnace: Rapid heating (0–1.000°C in 20 minutes) with silicone carbide elements.
NiCr-Ni Thermocouple: ±0.5°C accuracy across 300–1.000°C range.
Gas Control:
Mass flow controllers for nitrogen (1–5 L/min) and air (0.5–2 L/min).
Built-in leak detector with audible/visual alerts.
Weighing Module:
Integrated analytical balance (0.01mg resolution) with anti-vibration table.
Auto-tare function for crucibles and samples.
Data Management:
100-test memory with date/time stamping.
USB/RS-232 ports for PC connectivity and LIMS integration.
Safety Suite:
Emergency off button with E-Stop loop monitoring.
Fume extraction port for lab ventilation systems.
Technical Parameters
| Specification | Details |
|---|---|
| Temperature Range | 300°C to 1.000°C (max. operating temp.) |
| Temperature Accuracy | ±0.5°C (±0.1°C with calibration kit) |
| Heating Rate | 0–50°C/min (adjustable) |
| Gas Purity | 99.99% N₂ (inert) / Ambient air (combustion) |
| Sample Capacity | 1–5 samples per run (10g max. per sample) |
| Power Supply | 220V/50Hz (110V/60Hz available) |
| Dimensions (W×D×H) | 500mm × 400mm × 450mm |
| Weight | 45 kg (net) |